SF-479 medical freezer temperature humidity controller


Product number:1033013872803274752

  • Description
  • Features 

    Humidity display

    Temperature control

    Defrost control

    Light control / Power switch

    Value storing / Self-testing

     Evap. Fan control

     

    Technical Data 

    Temperature display range: -45~99℃

    Set temperature range: -45~+75℃

    Accuracy: ±01℃        

    Temperature sensor: NTC, one sensor(neither positive nor negative)2m(L)

    Compressor relay : 30A/250VAC

    Heater:  N.O. 10A/250VAC

    Others:  N.O. 5A/250VAC

    Dimension: 181mm( )×41mm(W)×68mm(D)

    Mounting hole dimension: 138mm( )×32.5mm(W)


    Parameters  

    Paramete

    Function

    Set range

    Default

    PA

    Menu password

    00~50~99

    82

    E1

    Lower set point limit

    -45℃~retain freshness set temp.

    00℃

    E2

    Higher set point limit

    retain freshness

    set temp.~75℃

    30℃

    E3

    Refrigeration Hysteresis

    01~20℃

    04℃

    E4

    Heating hysteresis

    01~20℃

    06℃

    E5

    Refrigeration start delay time

    00~10Min

    02Min

    E6

    Heating start delay time 

    00~10Min

    00Min

    E7

    Refrigeration and heating

    00~90Min

    15Min

    E8

    Offset on room sensor

    -20~20℃

    00℃

    E9

    Offset on humidity sensor

    -20~20%

    00%

    P1

    Light duration time 

    00~180 min

    00 is infinite length

    00min

    C1

    High temp. alarmvalue

    C2~75℃

    45℃

    C2

    low temp. alarmvalue

    -45℃~C1

    -35℃

    C3

    Alarm delay time

    00~60min

    2min

    C4

    Alarm hysteresis

    01~10℃

    02℃

    F1

    Defrost duration

    01~90min

    20min

    F2

    Defrost interval

    00~24hours

    4hr

    F4

    Display when defrost

    00=normal display

    01=locked display

    00

     

     

     

     

     

    H1

     

    Fan control

    H1=01synchronized with heater and comp., off  when defrost

    H1=02synchronized with heater and comp., open when defrost

    H1=03 keep running, open when defrost

    H1=04  keep running, close when defrost

    82